Stago STart 4 Analyzer Plugin

Feature: 011-madagascar-analyzer-integration
Milestone: M11 (Stago STart 4)
Status: ✅ Complete

Overview

External plugin JAR for Stago STart 4 coagulation/hemostasis analyzer. Supports dual-protocol communication:

  • ASTM LIS2-A2 over RS232 serial connection
  • HL7 v2.5 over Network (TCP/IP)

Analyzer Specifications

  • Manufacturer: Stago
  • Model: STart 4
  • Type: Coagulation/Hemostasis Analyzer
  • Protocols: ASTM LIS2-A2 (RS232), HL7 v2.5 (Network)
  • Tests: PT, INR, APTT, Fibrinogen, TT

Protocol Support

ASTM Format

The plugin automatically detects ASTM messages by checking for H-segment headers containing "START4" or "STAGO".

Message Structure:

  • H (Header) - Analyzer identification
  • P (Patient) - Patient demographics
  • O (Order) - Test order information
  • R (Result) - Coagulation test results
  • L (Terminator) - Message end

Example R-segment: R|1|^^^PT|12.5|sec|11.0-13.5|N||F|20260128080100

HL7 Format

The plugin automatically detects HL7 messages by checking for MSH segments with sending application "STAGO".

Message Structure:

  • MSH (Message Header) - Message metadata
  • PID (Patient Identification) - Patient demographics
  • ORC (Common Order) - Order information
  • OBR (Observation Request) - Test order details
  • OBX (Observation/Result) - Coagulation test results

Example OBX segment: OBX|1|NM|^^^PT^PROTHROMBIN TIME||12.5|sec|||||F||||||

Test Mappings

The plugin maps the following coagulation tests to OpenELIS via LOINC codes:

Analyzer Code OpenELIS Test Name LOINC Code
PT Prothrombin Time 5902-2
INR International Normalized Ratio 6301-6
APTT Activated Partial Thromboplastin Time 3173-2
FIB Fibrinogen 3255-7
TT Thrombin Time 3174-0

Integration with MappingAware

The plugin automatically integrates with the MappingAware wrapper pattern (Feature 004). When field mappings are configured for the Stago STart 4 analyzer in the OpenELIS dashboard, the system automatically wraps the plugin inserter with MappingAwareAnalyzerLineInserter to apply field mappings before processing results.

No plugin code changes required - integration is handled automatically by ASTMAnalyzerReader and HL7AnalyzerReader.

Installation

  1. Build the plugin:

    cd plugins/analyzers/StagoSTart4
    mvn clean install
  2. Deploy the JAR:

    cp target/StagoSTart4-1.0.jar /var/lib/openelis-global/plugins/
  3. Restart OpenELIS to load the plugin.

Testing

Unit tests are located in:

  • src/test/java/uw/edu/itech/StagoSTart4/StagoSTart4AnalyzerLineInserterTest.java

Test fixtures are located in the main OpenELIS project:

  • src/test/resources/testdata/stago/stago-start4-coagulation.astm
  • src/test/resources/testdata/stago/stago-start4-coagulation.hl7

References

  • Feature Specification: specs/011-madagascar-analyzer-integration/spec.md
  • Research Documentation: specs/011-madagascar-analyzer-integration/research.md
  • Task Breakdown: specs/011-madagascar-analyzer-integration/tasks.md

License

Mozilla Public License Version 1.1 (MPL 1.1)

Copyright (C) CIRG, University of Washington, Seattle WA. All Rights Reserved.